A Comparative Study of Intraocular Pressure Measurements Using Goldmann Applanation Tonometer, I-care pro and Corvis ST after DSAEK

Publish Year: 1397
نوع سند: مقاله کنفرانسی
زبان: English
View: 519

نسخه کامل این Paper ارائه نشده است و در دسترس نمی باشد

  • Certificate
  • من نویسنده این مقاله هستم

استخراج به نرم افزارهای پژوهشی:

لینک ثابت به این Paper:

شناسه ملی سند علمی:

ACSOMED28_097

تاریخ نمایه سازی: 1 دی 1397

Abstract:

Purpose: To compare the intraocular pressure (IOP) measurements using different devices and evaluate the influence of central corneal thickness (CCT) on IOP measurements following DSAEK surgery. Methods: In this cross-sectional study 22 eyes from 19 patients (mean age: 56.59 years) after DSAEK with clear and compact graft were enrolled. Mean follow-up after the surgery was 2.191±1.852 years. IOP was measured with GAT, I-care pro and Corvis ST (as non-corrected and biomechanical IOP) in post-DSAEK patients and agreement between measurements of three devises were assessed. Corvis ST was used to measure the CCT value. The influence of CCT on IOP measurements were evaluated using spearman test. Results: The mean IOP measured by GAT, I-care pro and Corvis ST the non-corrected and biomechanical IOP were 12.545±2.064, 11.591±2.465, 12.686±3.967 and 9.714±4.339 respectively. GAT measurement was considered as a gold standard and the only significant difference was observed between GAT and Corvis biomechanical IOP (P<0.05). CCT did not correlate with any of the IOP measurements except Corvis biomechanical IOP (r= -0.518).Conclusion: There was a good agreement between GAT, I care and Corvis non-corrected IOP measurements while the Corvis biomechanical IOP measurement was less than the GAT.

Authors

Shiva Bagheri

Vision Health Research Center, Semnan University of Medical Sciences, Semnan, Iran

Khosrow Jadidi

Vision Health Research Center, Semnan University of Medical Sciences, Semnan, Iran

Shiva Pirhadi

Vision Health Research Center, Semnan University of Medical Sciences, Semnan, Iran

Farhad Nejat

Vision Health Research Center, Semnan University of Medical Sciences, Semnan, Iran