Abstract:
Mechanical heart valves (MHVs) are widely implanted to replace the diseased heart valves. The performance of these valves strongly is limited by complications such as hemolysis and platelet activation. Abnormal flow patterns and turbulence generated by MHVs are the major reasons of these problems. Thus a better understanding of MHV hemodynamics is very important. In the present study, a simplified analytical approach is proposed to predict turbulent shear-induced hemolysis and platelet activation downstream of mechanical hear valves. This theoretical model was validated with such available in vivo experimental results where a good agreement was found. Therefore the present model provides a new insight into the behavior of turbulent Reynolds stresses downstream of the various mechanical heart valves and estimates them accurately without any extra computational or experimental cost.
Authors
مراجع و منابع این Paper:
لیست زیر مراجع و منابع استفاده شده در این Paper را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود Paper لینک شده اند :