Abstract:

This study aims to validate, using finite element analysis (FEA), the design concept by comparing the fatigue behavior of hip implant stems coated with composite (carbon/PEEK) and polymeric (PEEK) coating materials corresponding to different human activities: standing up, normal walking and climbing stairs under dynamic loadings to find out which of all these models have a better performance in the prosthesis-bone systems. A 3D finite element models of hip implants, femur, coating layers with polymeric (PEEK) and composite (carbon/PEEK) coating materials are created for FEA. The cyclic loads are applied on the prosthesis head. Fatigue life durations are calculated based on the Goodman mean-stress fatigue theory. The fatigue safety factor for the coated implant is increased more than 12.73% at least compared to the uncoated implant. The carbon/PEEK composite material with 0, +45, -45, and 90 degrees fiber orientation (configuration I) has the highest fatigue life and fatigue safety factor. The numerical result show that the carbon/PEEK composite material (configuration I) seems to be a good solution to increase the values of fatigue safety factor of coating layers due to highest fatigue life and fatigue safety factor. It distributes the applied load and transfers it to the bone, reducing stress-shielding effects and prolong the bone-prosthesis system life span.

Authors

Ayham Darwich

Faculty of Biomedical Engineering, Al-Andalus University for Medical Sciences, Tartous, Syria

Hasan Nazha

Faculty of Technical Engineering, University of Tartous, Tartous, Syria

Monzer Daoud

Department of Mechanical Engineering, École de Technologie Supérieure (ÉTS), Montreal, Canada

مراجع و منابع این Paper:

لیست زیر مراجع و منابع استفاده شده در این Paper را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود Paper لینک شده اند :
  • Holzwarth, U., Cotogno, G., Total Hip Arthroplasty: State of the ...
  • Petrolo, L., Testi, D., Taddei, F., Viceconti, M., Effect of ...
  • Scifert, C. E., A finite element investigation into the biomechanics ...
  • Hernandez-Rodriguez, M. A. L., Ortega-Saenz, J. A., Contreras-Hernandez, G. R., ...
  • Lin, Y. T., Wu, J. S. S., Chen, J. H., ...
  • Khanuja, H. S., Vakil, J. J., Goddard, M. S., & ...
  • John, A. A., Jaganathan, S. K., Supriyanto, E., Manikandan, A., ...
  • Jemat, A., Ghazali, M. J., Razali, M., Otsuka, Y., Surface ...
  • Hedia, H. S., Fouda, N., A new design of dental ...
  • Asiri, S. A., Hedia, H. S., Fouda, N., Improving the ...
  • Aldousari, S. M., Fouda, N., Hedia, H. S., AlThobiani, F. ...
  • Apostu, D., Lucaciu, O., Berce, C., Lucaciu, D., Cosma, D., ...
  • Branemark, P. I., Zarb, G., Albrektsson, T., Tissue- integrated prostheses: ...
  • Anguiano-Sanchez, J., Martinez-Romero, O., Siller, H. R., Diaz-Elizondo, J. A., ...
  • Liang, J., Kalyanasundaram, S., Effect of fiber orientation on the ...
  • Kaddick, C., Stur, S., Hipp, E., Mechanical simulation of composite ...
  • Kayabasi, O., Ekici, B., The effects of static, dynamic and ...
  • Stolk, J., Verdonschot, N., Murphy, B. P., Prendergast, P. J., ...
  • Han, X., Yang, D., Yang, C., Spintzyk, S., Scheideler, L., ...
  • Fujihara, K., Teo, K., Gopal, R., Loh, P. L., Ganesh, ...
  • Colombi, P., Fatigue analysis of cemented hip prosthesis: damage accumulation ...
  • Rezaei, F., Hassani, K., Solhjoei, N., Karimi, A., Carbon/PEEK composite ...
  • Caouette, C., Yahia, L. H., Bureau, M. N., Reduced stress ...
  • Budynas, R. G., Nisbett, K. J., Shigley’s mechanical engineering design. ...
  • Bergmann, G., Bender, A., Dymke, J., Duda, G., Damm, P., ...
  • Darwich, A., Nazha, H., & Abbas, W., Numerical study of ...
  • Ataollahi Oshkour, A., Talebi, H., Shirazi, S., Farid, S., Bayat, ...
  • Williams, P. L., Warwick, R., Dyson, M., Bannister, L. H., ...
  • Kayabasi, O., Ekici, B., The effects of static, dynamic and ...
  • Sobieraj, M. C., Murphy, J. E., Brinkman, J. G., Kurtz, ...
  • Al-Hmouz, I. A., The effect of loading frequency and loading ...
  • Uda, N., Ono, K., Kunoo, K., Compression fatigue failure of ...
  • Oshkour, A. A., Abu Osman, N. A., Yau, Y. H., ...
  • Herrmann, S., Kaehler, M., Souffrant, R., Rachholz, R., Zierath, J., ...
  • Brockett, C. L., Carbone, S., Abdelgaied, A., Fisher, J., Jennings, ...
  • Prochor, P., Finite element analysis of stresses generated in cortical ...
  • Hedia, H. S., & Fouda, N., Design optimization of cementless ...
  • Hedia, H. S., & Fouda, N., Improved Stress Shielding of ...
  • Enab, T., Behavior of FGM-coated, HA-coated and uncoated femoral prostheses ...
  • Fouda, N., Horizontal functionally graded material coating of cementless hip ...
  • Hedia, H. S., & Fouda, N. (2015). Improve the performance ...
  • نمایش کامل مراجع