Comparison of two automated methods for the measurement of glycated hemoglobin HBA۱C: ADAMS A۱c (ARKRAY) versus Capillarys ۲ Flex Piercing® (SEBIA)
عنوان مقاله: Comparison of two automated methods for the measurement of glycated hemoglobin HBA۱C: ADAMS A۱c (ARKRAY) versus Capillarys ۲ Flex Piercing® (SEBIA)
شناسه ملی مقاله: JR_MML-4-2_005
منتشر شده در در سال 1400
شناسه ملی مقاله: JR_MML-4-2_005
منتشر شده در در سال 1400
مشخصات نویسندگان مقاله:
Noussaiba Benhamza - Faculty of Medicine and Pharmacy of Oujda, Mohammed First University, Oujda, Morocco.
El-houcine Sebbar - Faculty of Medicine and Pharmacy of Oujda, Mohammed First University, Oujda, Morocco.
Adnane Naili - Faculty of Medicine and Pharmacy of Oujda, Mohammed First University, Oujda, Morocco.
Mohammed Choukri - Faculty of Medicine and Pharmacy of Oujda, Mohammed First University, Oujda, Morocco.
خلاصه مقاله:
Noussaiba Benhamza - Faculty of Medicine and Pharmacy of Oujda, Mohammed First University, Oujda, Morocco.
El-houcine Sebbar - Faculty of Medicine and Pharmacy of Oujda, Mohammed First University, Oujda, Morocco.
Adnane Naili - Faculty of Medicine and Pharmacy of Oujda, Mohammed First University, Oujda, Morocco.
Mohammed Choukri - Faculty of Medicine and Pharmacy of Oujda, Mohammed First University, Oujda, Morocco.
Introduction: The aim of this work is to present the results of a comparative study between the determination of glycated hemoglobin (HBA۱C) on the ADAMS A۱c® (ARKRAY) and Capillarys ۲ Flex Piercing® (SEBIA).
Materials and methods: ۳۱۰ venous blood samples were randomly selected from routine HBA۱C tests. The HBA۱C assay was performed on ADAMS A۱c® (ARKRAY) and Capillarys ۲ Flex Piercing® (SEBIA) during the same day. The data obtained were analyzed by the statistical software MedCalc Version ۱۵.۱.۰.
Results: The results obtained show a good correlation between the ۲ methods: the equation of the Passing-Bablok line is of type Y (Capillarys ۲ Flex Piercing®) = -۰.۵۵۰ + ۱.۱۱۹ X (ADAMS A۱c) the ۹۵% confidence interval of this slope is -۰.۶۴۶۷ to - ۰.۴۴۱۴ with r = ۰.۹۸۲ and p <۰.۰۰۰۱. The Bland-Altman plot shows that the average bias between the two methods is in the order of ۰.۳ and that the difference between the Capillarys and HPLC measurements of Hba۱c is in the range of +۱.۹۶ to -۱.۹۶ and the Deming regression equation Y (Capillarys ۲ Flex Piercing®) = -۰.۳۳۸۸ + ۱.۰۹۱۱ X (ADAMS A۱c).
Conclusion: Our study shows a good agreement of HBA۱C results between the Capillarys ۲ Flex Piercing® (SEBIA) and ADAMS A۱c® (ARKRAY). Laboratory work requires professionals to take into account variations in results when changing methods in their routine work by comparing their results on the various methods.
کلمات کلیدی: Glycatedhaemoglobin, Assay, Comparison, ADAMSA۱c®(ARKRAY), Capillarys۲FlexPiercing®(SEBIA)