A three-year survey of film badge service performance in Iran

عنوان مقاله: A three-year survey of film badge service performance in Iran
شناسه ملی مقاله: JR_IRJNM-3-1_003
منتشر شده در در سال 1374
مشخصات نویسندگان مقاله:

H Mohammadi - Research Institute for Nuclear Medicine, Tehran University of Medical Sciences, Tehran, Iran
S Mehdizadeh - Radiation Center, Shiraz University, Shiraz, Iran
Sh Derakhshan - Radiation Center, Shiraz University, Shiraz, Iran

خلاصه مقاله:
An extensive study was undertaken to Investigate on the overall film badge service performance with which occupational radiation, exposures are reported In Iran, The study involved the analysis of film badges exposed to x-and gamma rays and beta particles. For test exposures, x-rays between ۶۰ and ۱۴۰ KVPCo-۶۰ gamma radiation and Sr-۹۰ beta particles were used. The actual delivered dose equivalents were limited to ۰.۴۰-۵ mSv which is the range of annual occupational dose reported by many international studies. This study revealed that the reporting on gamma rays, specially for Co-۶۰, was very good. The reports on x-rays were overestimated. The overestimation was considerable at higher kVp's. The results for beta particles were rather poor.

کلمات کلیدی:
Low level radiation, Film badge dosimetry, Occupational radiation exposure