Evaluation and Comparison of Dosimetric Characteristics of Semiflex®۳D and Microdiamond in Relative Dosimetry under ۶ and ۱۵ MV Photon Beams in Small Fields

عنوان مقاله: Evaluation and Comparison of Dosimetric Characteristics of Semiflex®۳D and Microdiamond in Relative Dosimetry under ۶ and ۱۵ MV Photon Beams in Small Fields
شناسه ملی مقاله: JR_JBPE-12-5_005
منتشر شده در در سال 1401
مشخصات نویسندگان مقاله:

- - - MSc, Department of Medical Physics, School of Medicine, Shahid Sadoughi University of Medical Sciences, Yazd, Iran
- - - MD, Department of Radiation Oncology, School of Medicine, Kerman University of Medical Sciences, Kerman, Iran
- - - MD, Yazd Diabetes Research Center, Shahid Sadoughi University of Medical Sciences, Yazd, Iran
- - - PhD, Department of Medical Physics, School of Medicine, Shahid Sadoughi University of Medical Sciences, Yazd, Iran

خلاصه مقاله:
Background: In modern radiotherapy techniques, the frequently small and non-uniformed fields can increase treatment efficiency due to their highly conformal dose distribution. Particular features including lack of Lateral Charge Particle Equilibrium (LCPE) lead to detectors with high resolution since any error in obtained dosimetric data could cause patient mistreatments.Objective: This study aims to evaluate and compare two small detectors (Semiflex®۳D and microdiamond) dosimetric characteristics in small field relative dosimetry.Material and Methods: In this experimental study, the dosimetric properties of Semiflex®۳D and microdiamond were assessed under ۶ and ۱۵ MV photon beams. The linearity and stability of the detector’s response and dose rate were measured. Square-field sizes ranging from ۰.۶×۰.۶ - ۵×۵ cm۲ were used for obtaining percentage depth dose curves (PDDs) and in-plane profiles. The angular and temperature dependence of both detectors’ responses were also studied.Results: The detector response shows good stability, no deviation from linearity, and low dose rate dependence (≤۱.۶%). PDDs and in-plan profiles of both detectors are in good agreement and no significant difference was observed except for the high dose gradient regions (P-value≤۰.۰۱۷). Both detectors demonstrated low angular dependence (<۰.۳%) with temperature dependence lower than ۱% for both detectors. Conclusion: The results indicate both investigated detectors were well performed in small field relative dosimetry and for measuring penumbra, it is better to use microdiamond detector.

کلمات کلیدی:
Radiotherapy, Conformal, Radiotherapy, High-Energy, Radiation Dosimeters, Small Field Dosimetry