Abstract:
Introduction: In the present scenario, high precision-radiotherapy is delivered through Linear accelerators in which the dose delivery is achieved by delivering the proper monitor unit (MU). Treatment planning for the patients is carried out through treatment planning systems (TPS) in which the precise computation of MU is crucial. This TPS - calculated MU has to be verified using manual calculations for accurate dose delivery. In this study, we incorporated our in-house developed multi leaf collimator(MLC) shaper software and the well-known Clarkson method to compare the calculated MUs to the TPS-generated MUs.Material and Methods: Conformal treatment plans of various sites of ۳۰ patients were randomly selected containing different MLC-shaped field sizes. All the fields were shaped using MLC (leaf width of ۱cm, ۴۰ pairs) in the TPS. MLC log files were exported and fed into the in-house shaper software to get crucial inputs for the Clarkson-based calculation. The Tissue Maximum Ratio(TMR) & Scatter Maximum ratio(SMR) were utilized in our investigation. The Clarkson MU calculation was compared with the TPS calculation method. Paired t-test was performed for the statistical significance.Results: The Clarkson method-based calculated had significant differences for all the esophageal cancers (p<۰.۰۵); however no significant difference was found in the other sites.Conclusion: The compared MUs were within the acceptable deviation with the TPS for Head & Neck, Prostrate and Cervical cancer. The estimated MUs had significant difference in non-homogenous medium. The shaper software can be further enhanced to receive MLC log files from the TPS.
Authors
Athiyaman M
Radiological Physics, Assistant Professor, S.P.Medical College,Bikaner, Rajasthan-۳۳۴۰۰۳
Hemalatha A
Assistant professor,department of radiological physics,SP Medical college
Mary Joan
Department of Radiological Physics, CMC Ludhiana, Punjab, India
ARUN CHOUGULE
DEPARTMENT OF RADIOLOGICAL PHYSICS, SMS MEDICAL COLLEGE RAJASTHAN UNIVERSITY OF HEALTH SCIENCES, JAIPUR
kumar HS
Department of Radiotherapy, SP Medical College, Bikaner, Rajasthan, India.
Gokul Raj
Sr.Demonstrator
مراجع و منابع این Paper:
لیست زیر مراجع و منابع استفاده شده در این Paper را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود Paper لینک شده اند :