Design and Evaluation of a Test Atmosphere System for Passive Samplers Development
Publish place: Archives of Occupational Health، Vol: 1، Issue: 1
Publish Year: 1396
نوع سند: مقاله ژورنالی
زبان: English
View: 444
This Paper With 6 Page And PDF Format Ready To Download
- Certificate
- من نویسنده این مقاله هستم
استخراج به نرم افزارهای پژوهشی:
شناسه ملی سند علمی:
JR_AOH-1-1_007
تاریخ نمایه سازی: 1 مرداد 1397
Abstract:
Background: Preparation of the test atmosphere is a crucial step in occupational health, toxicology, and analytical chemistry researches. Methods: A simple and efficient system was developed for preparation of standard test atmosphere for application in occupational and environmental exposure studies, especially passive sampler s validation. Results: The system applicability was tested in various temperature, humidity, and velocity conditions. Results showed that for testing the adsorbent samplers, the system should be in hybrid mode of dilution. Conclusion: The repeatability of the measurements in the proposed system was also examined that was in acceptable range. With double dilution system it is possible to produce concentrations less than part per million.
Keywords:
Authors
Abdulrahman Bahrami
Department of Occupational Health, Research Centre for Health Sciences, Faculty of Public Health, Hamedan University of Medical Sciences, Hamadan, Iran