The Dosimetric Effects of Different Multileaf Collimator Widths on Physical Dose Distributions
Publish place: Iranian Journal of Medical Physics، Vol: 15، Issue: 1
Publish Year: 1397
نوع سند: مقاله ژورنالی
زبان: English
View: 228
This Paper With 7 Page And PDF Format Ready To Download
- Certificate
- من نویسنده این مقاله هستم
استخراج به نرم افزارهای پژوهشی:
شناسه ملی سند علمی:
JR_IJMP-15-1_003
تاریخ نمایه سازی: 18 تیر 1398
Abstract:
Introduction: Geometric changes in the multileaf collimator (MLC) led to dosimetric considerations in intensity-modulated radiation therapy (IMRT) due to the number and size of the pixels in the intensity map, which are determined by the MLC leaf width. In this study, we evaluated the dosimetric effects of different MLC widths on physical dose distributions for IMRT plans.
Materials and Methods: Forty-two IMRT plans based on different MLC devices were generated and analyzed to study the effect of MLC width on plan quality.
Results: Improvements in IMRT plan quality using 0.4 cm leaf width in comparison with 1 cm leaf width were evaluated. The 0.4 cm leaf-based plans resulted in significantly higher Dmean, D98%, D95% , D5%, and V95 (58.86 Gy, 95.11%, 96.57%, 104%, and 97.92%, respectively) compared to the 1 cm leaf plans (58.66 Gy, 92.56%, 94.56%, 104.14%, and 95.72%, respectively). Conformation number (CN) for planning target volume in 0.4 cm leaf plans was significantly higher than the 1 cm leaf plans (0.74 vs. 0.67; P<0.05). In addition, the 0.4 cm leaf plans significantly improved dose homogeneity compared to the 1 cm leaf plans (1.08 vs. 1.10; P<0.05). We found that 0.4 cm leaf width significantly decreased the integral dose to normal tissue compared to the 1 cm leaf width (from 56.09 to 49.46 Gy.Kg P<0.05). Conclusion: No significant clinical differences were observed between the two plans for a serially functioning tissue, while the differences in mean doses were statistically significant for parallel functioning normal tissues.
Authors
Ehab Attalla
Prof. of Medical physics ; National Cancer Institute, cairo university & Children Cancer Hospital , Egypt
Ismail Eldesoky
Cairo University, Cairo, Egypt
مراجع و منابع این Paper:
لیست زیر مراجع و منابع استفاده شده در این Paper را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود Paper لینک شده اند :