Graphite Furnace Atomic Absorption Spectrometry for Biomonitoring of Blood Lead Level in Quality Control Blood Materials
Publish place: Journal of Advanced Biomedical Sciences، Vol: 13، Issue: 2
Publish Year: 1402
نوع سند: مقاله ژورنالی
زبان: English
View: 131
This Paper With 7 Page And PDF Format Ready To Download
- Certificate
- من نویسنده این مقاله هستم
استخراج به نرم افزارهای پژوهشی:
شناسه ملی سند علمی:
JR_JABS-13-2_005
تاریخ نمایه سازی: 31 تیر 1402
Abstract:
Background & Objective: Lead (Pb) is a major toxicant that has threatened human health for years. There is no safe level of exposure and deposition of a small amount of Pb in the human body has a negative impact on an individual's health. Despite recent reductions in its use, it has caused extensive environmental contamination as well. In the present study, the intralaboratory quality control assessment is reported as an initial step in lead determination by Graphite furnace atomic absorption spectrometry (GFAAS).
Materials & Methods: Varian SpectrAA-۲۲۰ with partition tube and deuterium background correction was used for the analysis. The evaluation quality control blood material was Seronorm trace elements whole blood (levels I &II).
Results: The lowest method limit of detection was ۰.۷۴ μg/dL, and based on the regression analysis, the lowest quantification concentration was ۳.۰۰ μg/dL (CV = ۴.۶ %). In addition, obtained %CV for Seronorm (I) & Seronorm (II) was ۷.۳% & ۵.۴% respectively.
Conclusion: In recent years, several diverse technological advancements in blood lead level determination have occurred. At the same time, demands for blood lead (BPb) level determination by GFAAS has been increased by public health in order to identify children exposed to lead in the environment, and by occupational health to reduce excessive exposure in the lead industries. The obtained results indicated that for BPb analysis, the method described by GFAAS is still a reference routine analytical technique that requires less time and no matrix matched standard curve, allowing blood samples to be run against aqueous calibration standards which eliminates the need for standard additions.
Keywords:
Lead Exposure- Graphite Furnace Atomic Absorption- Biomonitoring- Precision-Intralaboratory
Authors
مهری علی اصغرپور
Reference Health Laboratory, Ministry of Health & Medical Education, Tehran, Iran
مراجع و منابع این Paper:
لیست زیر مراجع و منابع استفاده شده در این Paper را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود Paper لینک شده اند :