Dose Distribution Evaluation and Independent Quality Check of Spherical INTRABEAM™ Applicators via Radiochromic EBT۲ Film Measurement
Publish place: Iranian Journal of Medical Physics، Vol: 16، Issue: 2
Publish Year: 1398
نوع سند: مقاله ژورنالی
زبان: English
View: 91
This Paper With 7 Page And PDF Format Ready To Download
- Certificate
- من نویسنده این مقاله هستم
استخراج به نرم افزارهای پژوهشی:
شناسه ملی سند علمی:
JR_IJMP-16-2_005
تاریخ نمایه سازی: 5 شهریور 1402
Abstract:
Introduction: The present study was conducted to implement a simple practical independent quality check of depth dose and isotropy of the Intrabeam™ therapeutic X-ray machine using radiochromic EBT۲ film. Material and Methods: Theindependent quality check of ۱.۵, ۳.۵, and ۵-cm spherical Intrabeam™ applicators was accomplished using particular EBT۲ film cutting pieces with internal rounded edges in a water phantom. Prior to this measure, the film was calibrated at three distances from the ۵-cm applicator in water to clarify the effects of beam spectrum and dose rate alteration on film response. To this end, three calibration curves were plotted. Results: The results of the one-way analysis of variance showed a critical difference between film pieces receiving equal doses at various distances (p <۰.۰۵). Therefore, depth dose curves were designed using all three calibration curves. Smaller applicators represented steeper dose fall-off, compared to the larger sizes. In this regard, ۱۴.۹۷%, ۱۷.۵۹%, and ۳۰.۹۲% of the relative mean doses were measured at ۱ cm depth of ۱.۵-cm, ۳.۵-cm, and ۵-cm applicators, respectively. A ۱۰%/۱mm gamma index was satisfied for the lateral dose evaluation of corresponding depth relative to Z-direction. Conclusion: The approach implemented in this study could be carried out as a rapid monthly quality check method for the dose distribution evaluation oftheIntrabeam device.
Keywords:
Authors
Omid Baziar
Medical Physics Department, Faculty of Medicine, Mashhad University of Medical Sciences, Mashhad, Iran
Hamid Gholamhosseinian
Medical Physics Research Centere, Faculty of medicine, Mashhad University of Medical Sciences, Mashhad, Iran
Kazem Anvari
Cancer Research Center, Mashhad University of Medical Sciences, Mashhad, Iran
مراجع و منابع این Paper:
لیست زیر مراجع و منابع استفاده شده در این Paper را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود Paper لینک شده اند :