AThree-Wavelength Pulse Oximeter for Carboxyhemoglobin Determination
Publish place: Journal of Iranian Association of Electrical and Electronics Engineers، Vol: 5، Issue: 2
Publish Year: 1387
نوع سند: مقاله ژورنالی
زبان: English
View: 482
This Paper With 7 Page And PDF Format Ready To Download
- Certificate
- من نویسنده این مقاله هستم
استخراج به نرم افزارهای پژوهشی:
شناسه ملی سند علمی:
JR_JIAE-5-2_001
تاریخ نمایه سازی: 13 تیر 1396
Abstract:
Pulse oximetry measures the relative concentration of arterial oxygenated hemoglobin and reduced hemoglobin. This paper presents the design of a pulse oximeter, which measures the presence of a third substance, carboxyhemoglobin (HbCO), by unfiltered LEDs at 660 nm, 810 nm, and 940 nm. Background information is given on the theory of pulse oximetry. Theoretical approximations and calculations which are implemented in our instrument are presented. Sensor design and a flowchart, showing the signal processing routine that accomplishes the required analysis is included. The problem of noise occurrence during the signal processing routine is also discussed.
Keywords:
Authors
R. P. Buinevicius
PhD Student, Department of Electrical and Computer Engineering University of Wisconsin ۱۴۱۵ Engineering Drive Madison, WI ۵۳۷۰۶ USA
J. G. Webster
Professor Emeritus, Department of Biomedical Engineering University of Wisconsin ۱۵۵۰ Engineering Drive Madison, WI ۵۳۷۰۶ USA