سیویلیکا را در شبکه های اجتماعی دنبال نمایید.

AThree-Wavelength Pulse Oximeter for Carboxyhemoglobin Determination

Publish Year: 1387
Type: Journal paper
Language: English
View: 515

This Paper With 7 Page And PDF Format Ready To Download

Export:

Link to this Paper:

Document National Code:

JR_JIAE-5-2_001

Index date: 4 July 2017

AThree-Wavelength Pulse Oximeter for Carboxyhemoglobin Determination abstract

Pulse oximetry measures the relative concentration of arterial oxygenated hemoglobin and reduced hemoglobin. This paper presents the design of a pulse oximeter, which measures the presence of a third substance, carboxyhemoglobin (HbCO), by unfiltered LEDs at 660 nm, 810 nm, and 940 nm. Background information is given on the theory of pulse oximetry. Theoretical approximations and calculations which are implemented in our instrument are presented. Sensor design and a flowchart, showing the signal processing routine that accomplishes the required analysis is included. The problem of noise occurrence during the signal processing routine is also discussed.

AThree-Wavelength Pulse Oximeter for Carboxyhemoglobin Determination Keywords:

AThree-Wavelength Pulse Oximeter for Carboxyhemoglobin Determination authors

R. P. Buinevicius

PhD Student, Department of Electrical and Computer Engineering University of Wisconsin ۱۴۱۵ Engineering Drive Madison, WI ۵۳۷۰۶ USA

J. G. Webster

Professor Emeritus, Department of Biomedical Engineering University of Wisconsin ۱۵۵۰ Engineering Drive Madison, WI ۵۳۷۰۶ USA