Introduction: Irreversible electroporation (IRE) is a process in which the membrane of the cancer cells are irreversibly damaged with the use of high-intensity electric pulses, which in turn leads to cell death. The IRE is a non-thermal way to ablate the cancer cells. This process relies on the distribution of the electric field, which affects the pulse amplitude, width, and electrical conductivity of the tissues. The present study aimed to investigate the relationship of the pulse width and intensity with the conductivity changes during the IRE using simulation.
Materials and Methods: For the purpose of the study, the COMSOL 5 software was utilized to predict the conductivity changes during the IRE. We used 4,000 bipolar and monopolar pulses with the frequency of 5 kHz and 1 Hz, width of 100 µs, and electric fields of low and high intensity. Subsequently, we built three-dimensional numerical models for the liver tissue.
Results: The results of our study revealed that the conductivity of tissue increased during the application of electrical pulses. Additionally, the conductivity changes increased with the elevation of the electric field intensity.
Conclusion: As the finding of this study indicated, the IRE with high-frequency and low electric field intensity could change the tissue conductivity. Therefore, the IRE was recommended to be applied with high frequency and low voltage.
Keywords:
Authors
amir khorasani
Dept. of Medical Physics, School of Medical Sciences, Tarbiat Modares University, Tehran, Iran
Seyed Mohamad Firoozabadi
Department of Medical Physics, School of Medical Sciences, Tarbiat Modares University, Tehran, Iran.
zeinab shankayi
Dept. of Medical Physics, School of Medical Sciences, Tarbiat Modares University, Tehran, Iran
مراجع و منابع این Paper:
لیست زیر مراجع و منابع استفاده شده در این Paper را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود Paper لینک شده اند :