A Comparative Study of the Construction of Positron Emission Tomography/Computed Tomography Facilities in Three South African Hospitals

Publish Year: 1399
نوع سند: مقاله ژورنالی
زبان: English
View: 162

This Paper With 9 Page And PDF Format Ready To Download

  • Certificate
  • من نویسنده این مقاله هستم

استخراج به نرم افزارهای پژوهشی:

لینک ثابت به این Paper:

شناسه ملی سند علمی:

JR_IJMP-17-2_003

تاریخ نمایه سازی: 1 مرداد 1399

Abstract:

Introduction: Development of higher energy modalities such as positron emission tomography/computed tomography (PET/CT), has led to more complex shielding problems. This is due to several factors, such as the radiopharmaceutical relatively high-administered activity, high patient throughput, and high energies of 511 kilo-electron volt (keV) positron annihilation photons. Therefore, this study aimed to compare three different methods used to determine the required shielding thicknesses of PET/CT facilities. Material and Methods: The required shielding thicknesses for three facilities were determined by using three different shielding methods, i.e. narrow beam, broad beam and Monte Carlo approximation. The design goal was chosen as 6 mSv/year for radiation workers and 1 mSv/year for the public. In addition, occupancy factors (T) were established, and all calculations had a use factor (U) of 1. The workload (W) of facilities and thicknesses of all barriers were then calculated for the three facilities. Results: For narrow beam approximation the average required thicknesses obtained were 6.16 mm lead, 5.12 cm concrete and 2.95 cm iron. Broad beam approximation required an average of 7.55 mm lead, 8.01 cm concrete and 2.96 cm iron thicknesses. Monte Carlo approximation required 7.62 mm lead, 10.59 cm concrete and 2.94 cm iron thicknesses. Conclusion: The narrow beam approximation demonstrated the least shielding thickness required for the materials used in this study, which can lead to under-shielding. The broad beam and Monte Carlo approximations demonstrated higher required shielding thickness although there were discrepancies between these two approximations for lead, concrete, and iron. 

Authors

Nompumelelo Masango

Department of Medical Physics, Faculty of Health Science, Sefako Makgatho Health Science University, Pretoria, South Africa

Bronwin Van Wyk

Department of Medical Physics, Faculty of Medicine, Dr George MUkhari Hospital, Pretoria, South Africa.

Modisenyane Mongane

University of free State Faculty: Health Sciences PO Box Bloemfontein ۹۳۰۰, Republic of South Africa