سیویلیکا را در شبکه های اجتماعی دنبال نمایید.

Optimization of Energy Window and Collimator for Y-90 Bremsstrahlung SPECT imaging: A Monte Carlo Simulation Study

Publish Year: 1400
Type: Journal paper
Language: English
View: 364

This Paper With 7 Page And PDF Format Ready To Download

Export:

Link to this Paper:

Document National Code:

JR_IJMP-18-4_005

Index date: 11 July 2021

Optimization of Energy Window and Collimator for Y-90 Bremsstrahlung SPECT imaging: A Monte Carlo Simulation Study abstract

Introduction: In yttrium-90 imaging, image quality is highly dependent on the selection of energy window and collimator design becausetheY-90 bremsstrahlung photons have a continuous and broad energy distribution. The current study aimed to optimize the bremsstrahlung energy window setting and collimator for the improvement of both resolution and sensitivity. Material and Methods: In the present study, simulation of medical imaging nuclear detectors (SIMIND) Monte Carlo program was used to simulate Siemens Medical System Symbia. The SIMIND was utilized to generate the Y-90 bremsstrahlung single-photon emission computed tomography (SPECT) projection of the point source. Six energy windows settings and two collimators denoting medium energy and high energy were used in order to assess the effect of the energy window on the resolution. Results: The experimental measurements and simulation results showed a similar pattern in the point spread functions with the energy window. The simulation data indicated that the geometric component reached 73%for the energy window within the range of51-120keVusingthe high-energy (HE) collimator. In addition, the obtained results showed that the full width at half maximum (FWHM) and full width at tenth maximum (FWTM)(FWHM=7mm and FWTM=35mm)were higher in this window in comparison to those reported for other windows. Conclusion: According to the obtained results of the present study, the optimal energy window for Y-90 bremsstrahlung SPECT imaging was within the range of 51-120 keV. The obtained optimal energy window and optimal HE collimator had the potential to improve the image resolution and sensitivity of Y-90 SPECT images

Optimization of Energy Window and Collimator for Y-90 Bremsstrahlung SPECT imaging: A Monte Carlo Simulation Study Keywords:

Y , 90 SPECT imaging Monte Carlo

Optimization of Energy Window and Collimator for Y-90 Bremsstrahlung SPECT imaging: A Monte Carlo Simulation Study authors

Hicham Asmi

Department of Physics, LPHE, Modeling and Simulations, Faculty of Science, Mohammed V University, Rabat, Morocco

Farida Bentayeb

Department of Physics,LPHE, Modeling and Simulations, Faculty of Science, Mohammed V University, Rabat, Morocco

Youssef Bouzekraoui

Department of Physics,LPHE, Modeling and Simulations, Faculty of Science, Mohammed V University, Rabat, Morocco

Faustino Bonutti

Academic Hospital of Udine, Department of Medical Physics, Udine, Italy

sanae douama

Department of Physics,LPHE, Modeling and Simulations, Faculty of Science,Mohammed V University, Rabat, Morocco

مراجع و منابع این Paper:

لیست زیر مراجع و منابع استفاده شده در این Paper را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود Paper لینک شده اند :
Stigbrand T, Carlsson J, Adams GP, editors. Targeted radionuclide tumor ...
Nag S, Kennedy A, Salem R, Murthy R, McEwan A, ...
Rong X, Frey EC. A collimator optimization method for quantitative ...
Rong X, Ghaly M, Frey EC. Optimization of energy window ...
Rong X, Du Y, Frey EC. A method for energy ...
Heard S, Flux GD, Guy MJ, Ott RJ. Photon source ...
Shen S, Denardo G, Yuan A, Denardo D and Denardo ...
Yue J, Mauxion T, Reyes DK, Lodge MA, Hobbs RF, ...
Elschot M, Vermolen B, Lam M, de Keizer B, van ...
Bouzekraoui Y, Bentayeb F, Asmi H, Bonutti F. Energy window ...
De Vries DJ, Moore SC. Approximation of hexagonal holes by ...
Ljungberg M. The SIMIND Monte Carlo program home page. ۲۰۱۸. ...
Ferreira T.Rasband W. Image J Program. ۲۰۱۸. Availablefrom:https://imagej.nih.gov/ij/download.html ...
نمایش کامل مراجع