Dosimetric Comparison of Two Linear Accelerator-Based Radiosurgery Systems for Intracranial Tumours with Rapidarc and Dynamic Conformal Arc Therapy

Publish Year: 1400
نوع سند: مقاله ژورنالی
زبان: English
View: 132

This Paper With 10 Page And PDF Format Ready To Download

  • Certificate
  • من نویسنده این مقاله هستم

استخراج به نرم افزارهای پژوهشی:

لینک ثابت به این Paper:

شناسه ملی سند علمی:

JR_IJMP-18-5_004

تاریخ نمایه سازی: 1 آبان 1400

Abstract:

Introduction: The present study focused on the dosimetric evaluation of Edge and Novalis Tx (NTx) linear accelerator (LA)-based radiosurgery system by using RapidArc (RA) and dynamic conformal arc (DCA) planning techniques. Material and Methods: Forty patients with brain lesions of variable sizes (۱.۱-۱۵.۹۸ cc) were planned for Edge and NTx system by using the RA and DCA planning techniques on eclipse treatment planning system, version ۱۳.۶ (Varian Medical Systems, Palo Alto, CA, USA). All the plans were evaluated on the basis of paddick conformity index (PCI), homogeneity index (HI), and gradient index (GI). The maximum doses to organs at risk (OAR), V۱۲Gy, V۱۰Gy, and V۵Gy for healthy brain tissue were also evaluated for all the plans. The treatment delivery efficiency for both systems was also evaluated. Results: The mean PCI and GI for both RA and DCA plans were found to be better in Edge as compared to NTx system (PCI Edge, RA=۰.۷۷±۰.۱, PCI NTx, RA=۰.۶۶±۰.۱۱, PCI Edge, DCA= ۰.۶۹±۰.۱۲, PCI NTx,DCA= ۰.۶۷±۰.۱۲). Significant differences in HI, doses to OAR, and V۱۲Gy, V۱۰Gy, and V۵Gy brain volume were observed for both systems with p-value less than ۰.۰۵. Reduced treatment time was observed in Edge LA as compared to NTx LA. Conclusion: Edge LA produced clinically better target volume conformity, rapid dose fall-off, and reduced reduction in normal brain volume irradiation and treatment time compared to NTx. Thus, in the set of patient plans evaluated, it was noted that Edge stereotactic suite is more efficacious and diametrically suitable for intracranial radiosurgery.   

Authors

Vaibhav Mhatre

Department of Radiation Oncology and Neurosurgery, Kokilaben Dhirubhai Ambani Hospital and Medical Research Institute, Andheri, Mumbai, India.

Pranav Chadha

Department of Radiation Oncology and Neurosurgery, Kokilaben Dhirubhai Ambani Hospital and Medical Research Institute, Andheri, Mumbai, India.

Rajkumar Chauhan

Department of Radiation Oncology and Neurosurgery, Kokilaben Dhirubhai Ambani Hospital and Medical Research Institute, Andheri, Mumbai, India.

Kaustav Talapatra

Department of Radiation Oncology and Neurosurgery, Kokilaben Dhirubhai Ambani Hospital and Medical Research Institute, Andheri, Mumbai, India.

Abhaya Kumar

Department of Radiation Oncology and Neurosurgery, Kokilaben Dhirubhai Ambani Hospital and Medical Research Institute, Andheri, Mumbai, India.

مراجع و منابع این Paper:

لیست زیر مراجع و منابع استفاده شده در این Paper را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود Paper لینک شده اند :
  • Wu A. Physics and dosimetry of the gamma knife. Neurosurg ...
  • Gevaert T, Levivier M, Lacornerie T, Verellen D, Engels B, ...
  • Lagerwaard FJ, Meijer O, Hoorn VE, Verbakel WF, Slotman BJ, ...
  • Kim Ji, Park JM, Park SY, Choi CH, Wu HG, ...
  • Joy S, Starkschall G, Kry S, Salehpour M, White RA, ...
  • Prendergast BM, Popple RA, Clark GM, Spencer SA, Guthrie B, ...
  • Thomas EM, Popple RA, Wu X,Clark GM, Markert JM, Guthrie ...
  • Ruschin M, Sahgal A, Iradji S, Soliman H, Leavens C, ...
  • نمایش کامل مراجع