Highly Selective Perchlorate Coated-Wire Electrode (CWE) based on an Electrosynthesized Dixanthylinum Dye and Its Application in Water Samples

Publish Year: 1401
نوع سند: مقاله ژورنالی
زبان: English
View: 160

This Paper With 9 Page And PDF Format Ready To Download

  • Certificate
  • من نویسنده این مقاله هستم

استخراج به نرم افزارهای پژوهشی:

لینک ثابت به این Paper:

شناسه ملی سند علمی:

JR_CHRL-5-4_004

تاریخ نمایه سازی: 12 شهریور 1401

Abstract:

The fabrication, performance, and utilization of perchlorate coated-wire electrode (perchlorate-CWE) based on dixanthylinum dye concerning the response towards perchlorate ions was defined. The recommended electrode exhibits a Nernstian response for perchlorate by a wide concentrations range of ۱.۰×۱۰−۶ to ۶.۱×۱۰−۲ M, with a slope of -۵۷.۴ mV in comparison to other anions. The limit of detection is ۵.۰×۱۰−۷ M. It has a quick response time of ~۴ seconds. The recommended electrode shows adequately good discriminating ability towards perchlorate ions in comparison to other anions and it is profitably applied to the direct determination of perchlorate ions in water samples.

Keywords:

Coated wire ion-selective electrode , Potentiometers , Perchlorate , Dixanthylinum , Perchlorate-CWE