Abstract:
Introduction:A sleep apnea monitor (BM۲۰۰۰A) is a wrist-worn device that measures oxygen saturation and pulse rate during sleep. This study aimed to evaluate the efficacy of the watch-like BM۲۰۰۰A for screening obstructive sleep apnea (OSA).BM۲۰۰۰A; Home sleep apnea testing; Obstructive sleep apnea; Polysomnography; Wrist pulse oximeter Materials and Methods:۱۰۲ patients complaining of sleep breathing disorders were included; ۸۱% were men and ۱۹% were women. All participants underwent overnight simultaneous polysomnography (PSG) and BM۲۰۰۰A sleep monitoring. The number of apneas and hypopneas, apnea-hypopnea index (AHI), percentage of time spent with oxygen saturation under ۹۰%, average oxygen saturation, lowest oxygen saturation, and duration of sleep were computed by the BM۲۰۰۰A and PSG. Then, these parameters were compared to validate the BM۲۰۰۰A.Results:All parameters, measured with BM۲۰۰۰A, had a good correlation (r ≥ ۰.۶, p < ۰.۰۰۰۱) with PSG-derived indexes, except for sleep time (r = ۰.۱۹, p = ۰.۰۶۱) and hypopnea index (r = ۰.۴, p < ۰.۰۰۰۱). AHI had the strongest correlation (r = ۰.۸۷, p < ۰.۰۰۰۱). The mean difference between AHI values calculated with PSG and wrist-worn pulse oximeter (WPO) was -۱۷.۶۶ events/h (۹۵% CI: -۵۰.۳۹ to ۱۵.۰۶). In AHI ≥ ۵, BM۲۰۰۰A had ۹۰.۷% sensitivity, ۱۰۰% specificity, ۹۱.۲% accuracy, and ۰.۹۹۴ area under the curve. Using AHI ≥ ۵, ≥ ۱۵, and ≥۳۰ as the screening criteria, optimal WPO-AHI cutoffs to improve the screening accuracy were ۳.۱۰, ۸.۹۲, and ۱۳.۰۵.Conclusions:BM۲۰۰۰A-derived results properly correlate with PSG and can provide OSA screening with good sensitivity and specificity.
Keywords:
BM۲۰۰۰A , Home sleep apnea testing , Obstructive Sleep Apnea , Polysomnography , Wrist pulse oximeter
Authors
Sama Jabbaripour
Medical Student, School of Medicine, Tehran University of Medical Sciences, Tehran, Iran.
Sareh Saien
Medical Student, School of Medicine, Tehran University of Medical Sciences, Tehran, Iran.
Reihaneh Heidari
Otolaryngology Research Center, Vali Asr Hospital, Imam Khomeini Hospital Complex, Tehran University of Medical Sciences, Tehran, Iran.
Reza Erfanian
Otolaryngology Research Center, AmirAlam Hospital, Tehran University of Medical Sciences, Tehran, Iran.
Amin Amali
Otolaryngology Research Center, Vali Asr Hospital, Imam Khomeini Hospital Complex, Tehran University of Medical Sciences, Tehran, Iran.
مراجع و منابع این Paper:
لیست زیر مراجع و منابع استفاده شده در این Paper را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود Paper لینک شده اند :