Validation and evaluation of a GATE model for MAMMI PET scanner

Publish Year: 1399
نوع سند: مقاله ژورنالی
زبان: English
View: 430

This Paper With 6 Page And PDF Format Ready To Download

  • Certificate
  • من نویسنده این مقاله هستم

استخراج به نرم افزارهای پژوهشی:

لینک ثابت به این Paper:

شناسه ملی سند علمی:

JR_IRJNM-28-1_005

تاریخ نمایه سازی: 19 بهمن 1398

Abstract:

Introduction:MAMMI is a dedicated PET based on high resolution detectors placed close to the breast. In this study, we presented a GATE model for the simulation of MAMMI scanner and model its performance of the MAMMI based on an adaptation of the NU 4-2008 NEMA standard.  Methods:A detailed of geometry MAMMI system that uses scintillation crystals coupled to position sensitive photomultipliers. The detector ring consists of 12 LYSO detector modules with a scanner aperture of 186 mm. We validated the model against experimental measurement, including spatial resolution, sensitivity, counting rates, noise equivalent count rate (NECR) and scattering pattern. Results: Overall results showed reasonable agreement between simulation and experimental data. For a breast phantom with a capillary source, the SF reaches 50.9 and NECR is 45 kcps with an activity of 11 MBq, 18F in quad rings. The spatial resolution at the axial FOV of quad rings (2.1 mm axial, 1.8 mm radial, and 1.7 tangential) is slightly better than that measured at the axial center of dual rings (2.1 mm axial, 1.8 mm tangential and radial). Conclusion: The MAMMI-PET has excellent spatial resolution and high sensitivity for primary breast cancer lesions. The results show performance improvement, especially in the absolute sensitivity, because of the more rings introduced in the MAMMI PET. The performance of the scanner and the validation results are considered to be reasonable enough to support its use in breast cancer imaging.

Authors

Azadeh Emami

Research Center for Molecular and Cellular Imaging, Tehran University of Medical Sciences, Tehran, Iran|Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran, Iran|International Campus, Tehran

Pardis Ghafarian

Chronic Respiratory Diseases Research Center, National Research Institute of Tuberculosis and Lung Diseases, Shahid Beheshti University of Medical Sciences, Tehran, Iran|PET/CT and Cyclotron Center, Masih Daneshvari Hospital, Shahid Beheshti

Hossien Ghadiri

Research Center for Molecular and Cellular Imaging, Tehran University of Medical Sciences, Tehran, Iran|Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran, Iran

Parham Geramifar

Research Center for Nuclear Medicine, Tehran University of Medical Sciences, Tehran, Iran

مراجع و منابع این Paper:

لیست زیر مراجع و منابع استفاده شده در این Paper را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود Paper لینک شده اند :
  • Avril N, Rosé CA, Schelling M, Dose J, Kuhn W, ...
  • Thompson CJ, Murthy K, Weinberg IN, Mako F. Feasibility study ...
  • Moliner L, Gonzalez AJ, Soriano A, Sanchez F, Correcher C, ...
  • Jan S, Santin G, Strul D, Staelens S, Assié K, ...
  • Moliner L, Correcher C, Hellingman D, Alamo J, Carrilero V, ...
  • Santin G, Strul D, Lazaro D, Simon L, Krieguer M, ...
  • Kerhoas-Cavata S, Guez D. Modeling electronic processing in GATE. Nucl ...
  • Nikolopoulos D, Kottou S, Chatzisavvas N, Argyriou X, Vlamakis E, ...
  • Yan J, Chaal J, Majewski S, Vaigneur K, Townsend DW. ...
  • Levin CS, TaiY-C, Hoffman E, Dahlbom M, Farquhar T. Removal ...
  • Raylman RR, Abraham J, Hazard H, Koren C, Filburn S, ...
  • D ambrosio D, Fiacchi G, Marengo M, Boschi S, Fanti ...
  • Lee E, Werner ME, Karp JS, Surti S. Design optimization ...
  • نمایش کامل مراجع