Investigation of LiF, Mg and Ti (TLD-۱۰۰) Reproducibility

Publish Year: 1394
نوع سند: مقاله ژورنالی
زبان: English
View: 78

This Paper With 6 Page And PDF Format Ready To Download

  • Certificate
  • من نویسنده این مقاله هستم

استخراج به نرم افزارهای پژوهشی:

لینک ثابت به این Paper:

شناسه ملی سند علمی:

JR_JBPE-5-4_008

تاریخ نمایه سازی: 30 دی 1402

Abstract:

LiF, Mg and Ti cubical TLD chips (known as TLD-۱۰۰) are widely used for dosimetry purposes. The repeatability of TL dosimetry is investigated by exposing them to doses of (۸۱, ۱۶۲ and ۴۰.۵ mGy) with ۶۶۲keV photons of Cs-۱۳۷. A group of ۴۰ cubical TLD chips was randomly selected from a batch and the values of Element Correction Coefficient (ECC) were obtained ۴ times by irradiating them to doses of ۸۱ mGy (two times), ۱۶۲mGy and ۴۰.۵mGy. Results of this study indicate that the average reproducibility of ECC calculation for ۴۰ TLDs is ۱.۵%, while these values for all chips do not exceed ۵%.

Authors

M Sadeghi

Radiation Research Center, Shiraz University, Shiraz, Iran

S Sina

Nuclear Engineering Department, School of mechanical Engineering, Shiraz University, Shiraz, Iran

R Faghihi

Nuclear Engineering Department, School of mechanical Engineering, Shiraz University, Shiraz, Iran