Evaluation and Comparison of Dosimetric Characteristics of Semiflex®۳D and Microdiamond in Relative Dosimetry under ۶ and ۱۵ MV Photon Beams in Small Fields

Publish Year: 1401
نوع سند: مقاله ژورنالی
زبان: English
View: 56

This Paper With 12 Page And PDF Format Ready To Download

  • Certificate
  • من نویسنده این مقاله هستم

استخراج به نرم افزارهای پژوهشی:

لینک ثابت به این Paper:

شناسه ملی سند علمی:

JR_JBPE-12-5_005

تاریخ نمایه سازی: 30 دی 1402

Abstract:

Background: In modern radiotherapy techniques, the frequently small and non-uniformed fields can increase treatment efficiency due to their highly conformal dose distribution. Particular features including lack of Lateral Charge Particle Equilibrium (LCPE) lead to detectors with high resolution since any error in obtained dosimetric data could cause patient mistreatments.Objective: This study aims to evaluate and compare two small detectors (Semiflex®۳D and microdiamond) dosimetric characteristics in small field relative dosimetry.Material and Methods: In this experimental study, the dosimetric properties of Semiflex®۳D and microdiamond were assessed under ۶ and ۱۵ MV photon beams. The linearity and stability of the detector’s response and dose rate were measured. Square-field sizes ranging from ۰.۶×۰.۶ - ۵×۵ cm۲ were used for obtaining percentage depth dose curves (PDDs) and in-plane profiles. The angular and temperature dependence of both detectors’ responses were also studied.Results: The detector response shows good stability, no deviation from linearity, and low dose rate dependence (≤۱.۶%). PDDs and in-plan profiles of both detectors are in good agreement and no significant difference was observed except for the high dose gradient regions (P-value≤۰.۰۱۷). Both detectors demonstrated low angular dependence (<۰.۳%) with temperature dependence lower than ۱% for both detectors. Conclusion: The results indicate both investigated detectors were well performed in small field relative dosimetry and for measuring penumbra, it is better to use microdiamond detector.

Authors

- -

MSc, Department of Medical Physics, School of Medicine, Shahid Sadoughi University of Medical Sciences, Yazd, Iran

- -

MD, Department of Radiation Oncology, School of Medicine, Kerman University of Medical Sciences, Kerman, Iran

- -

MD, Yazd Diabetes Research Center, Shahid Sadoughi University of Medical Sciences, Yazd, Iran

- -

PhD, Department of Medical Physics, School of Medicine, Shahid Sadoughi University of Medical Sciences, Yazd, Iran

مراجع و منابع این Paper:

لیست زیر مراجع و منابع استفاده شده در این Paper را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود Paper لینک شده اند :
  • Würfel JU. Dose measurements in small fields. Med Phys. ۲۰۱۳;۱(۱):۸۱-۹۰ ...
  • Scott AJ, Kumar S, Nahum AE, Fenwick JD. Characterizing the ...
  • Das IJ, Ding GX, Ahnesjo A. Small fields: nonequilibrium radiation ...
  • Debnath SBC, Fauquet C, Tallet A, Gonçalves A, Lavandier S, ...
  • Poppinga D, Delfs B, Meyners J, Harder D, Poppe B, ...
  • Haryanto F, Fippel M, Laub W, Dohm O, Nüsslin F. ...
  • Palmans H, Andreo P, Huq MS, Seuntjens J, Christaki KE, ...
  • Dosimetry of Small Static Fields Used in External Beam Radiotherapy. ...
  • Almaviva S, Ciancaglioni I, Consorti R, De Notaristefani F, Manfredotti ...
  • Reggiori G, Mancosu P, Suchowerska N, Lobefalo F, Stravato A, ...
  • Marinelli M, Prestopino G, Verona C, Verona-Rinati G. Experimental determination ...
  • Ghazal M, Westermark M, Kaveckyte V, Carlsson-Tedgren Å, Benmakhlouf H. ...
  • De Coste V, Francescon P, Marinelli M, Masi L, Paganini ...
  • Lechner W, Palmans H, Sölkner L, Grochowska P, Georg D. ...
  • Smith CL, Montesari A, Oliver CP, Butler DJ. Evaluation of ...
  • Alfonso R, Andreo P, Capote R, Huq MS, Kilby W, ...
  • De Angelis C, Casati M, Bruzzi M, Onori S, Bucciolini ...
  • Biasi G, Al Shukaili K, Petasecca M, Corde S, Guatelli ...
  • Reggiori G, Stravato A, Pimpinella M, Lobefalo F, De Coste ...
  • Ralston A, Tyler M, Liu P, McKenzie D, Suchowerska N. ...
  • Ciancaglioni I, Marinelli M, Milani E, Prestopino G, Verona C, ...
  • Klenevskii A, Bogatov N. Effects of Ionizing Radiation Detector Characteristics ...
  • Looe HK, Büsing I, Tekin T, Brant A, Delfs B, ...
  • Keivan H, Shahbazi-Gahrouei D, Shanei A. Evaluation of dosimetric characteristics ...
  • The Dosimetry Company. Dosimetry Catolog [Internet]. Freiburg Germany: The Dosimetry ...
  • Denia PM, Castellet García MD, Manjón García C, et al. ...
  • Russo S, Reggiori G, Cagni E, Clemente S, Esposito M, ...
  • Fox C, Simon T, Simon B, Dempsey JF, Kahler D, ...
  • Laub WU, Wong T. The volume effect of detectors in ...
  • Cranmer-Sargison G, Weston S, Sidhu NP, Thwaites DI. Experimental small ...
  • Podgorsak EB. Radiation oncology physics: a handbook for teachers and ...
  • Andreo P, Burns D, Hohlfeld K, Huq MS, Kanai T, ...
  • Van Der Merwe D, Van Dyk J, Healy B, Zubizarreta ...
  • Lárraga-Gutiérrez JM, Ballesteros-Zebadúa P, Rodríguez-Ponce M, García-Garduño OA, De la ...
  • Brualla-González L, Gómez F, Pombar M, Pardo-Montero J. Dose rate ...
  • Andreo P. The physics of small megavoltage photon beam dosimetry. ...
  • Delfs B, Kapsch RP, Chofor N, Looe HK, Harder D, ...
  • Kampfer S, Cho N, Combs SE, Wilkens JJ. Dosimetric characterization ...
  • Islam NM, Ma T, Podgorsak MB. The impact of water ...
  • Akino Y, Gautam A, Coutinho L, Würfel J, Das IJ. ...
  • نمایش کامل مراجع