Characteristics of the Howland current source for Bioelectric Impedance Measurements Systems

Publish Year: 1392
نوع سند: مقاله کنفرانسی
زبان: English
View: 866

This Paper With 5 Page And PDF Format Ready To Download

  • Certificate
  • من نویسنده این مقاله هستم

استخراج به نرم افزارهای پژوهشی:

لینک ثابت به این Paper:

شناسه ملی سند علمی:

ICBME20_069

تاریخ نمایه سازی: 25 فروردین 1394

Abstract:

In the development of bioelectric impedance measuring systems, design of a stable current source with broad bandwidth and high output impedance is an important consideration.Howland circuit is a well known high performance, voltage controlled current source which can provide high outputimpedance as well as high frequency bandwidth. This paper presents a detailed study of Howland current source by considering the non-ideal characteristics of the components. Twodifferent forms of Howland current source are compared and resulted relationship were evaluated by simulations. The resultsof this study facilitate the design of Howland circuit for bioelectric impedance measurement systems

Authors

Hassan Yazdanian

Department of Biomedical Engineering University of Isfahan Isfahan, Iran

Mohsen Mosayebi Samani

Department of Biomedical Engineering Tarbiat Modares University Tehran, Iran

Amin Mahanm

Department of Biomedical Engineering University of Isfahan Isfahan, Iran