Determination of phosphate by carbon paste electrode modified with SMZ by potentiometric method
Publish place: Iran International Zeolite Conference
Publish Year: 1387
نوع سند: مقاله کنفرانسی
زبان: English
View: 1,548
This Paper With 6 Page And PDF Format Ready To Download
- Certificate
- من نویسنده این مقاله هستم
استخراج به نرم افزارهای پژوهشی:
شناسه ملی سند علمی:
ZEOLITE01_025
تاریخ نمایه سازی: 30 دی 1387
Abstract:
The utility of carbon paste electrode modified with SMZ (Na-Y zeolite modified with HDTMA
surfactant) for the potentiometric determination of phosphate in aqueous solution is demonstrated. The
electrode exhibits linear response to phosphate over a wide concentration range (1.58 × 10-5 – 1.00 × 10-2 M)
with Nernstian slope of 30 ± 2 mV per decade in temperature range 25 to 50 oC . It has a response time of
about 15 s and can be used for a period of two months with good reproducibility and repeatability. The
detection limit of this electrode was 1.5 × 10-5 M. The electrode response for phosphate determination in pH
range 6-12, was constant and independent from pH. This chemically modified carbon paste electrode was
successfully used for the determination of phosphate in ammonium phosphate chemically fertilizer sample.
Keywords:
مراجع و منابع این Paper:
لیست زیر مراجع و منابع استفاده شده در این Paper را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود Paper لینک شده اند :